The Tindeq Progressor has been validated in peer-reviewed research against laboratory reference instruments. Against a Kistler force plate in 25 sport climbers it agreed to ICC 0.99 (Labott et al., 2022). On the bench, against an Instron materials-testing machine, test–retest and validity ICC were ≥ 0.999 with errors below 0.9 N across 10–100 N (Merry et al., 2021). In 31 patients with knee disorders it tracked isokinetic dynamometry at least as well as hand-held dynamometry (Center et al., 2024), and across five hip and thigh actions in healthy adults reliability was ICC 0.93–0.99 (Jacques et al., 2025). Every study below is linked to its DOI so you can read the original.

Validation studies — climbing

Labott BK, Held S, Wiedenmann T, Rappelt L, Wicker P, Donath L (2022)

Validity and Reliability of a Commercial Force Sensor for the Measurement of Upper Body Strength in Sport Climbing. Frontiers in Sports and Active Living, 4:838358. doi:10.3389/fspor.2022.838358
n = 25 experienced climbers. Progressor (80 Hz) vs Kistler force plate (500 Hz): validity ICC 0.99 (95% CI 0.98–1.00) for one-handed rung pulling and bent-arm lock-off, SEM 0.9–1.7 N, CV 2.4–3.1%. Between-day reliability ICC 0.95–0.98.
Takeaway: near-perfect agreement with a laboratory force plate and excellent day-to-day reliability — the source of the “ICC 0.99” figure.

Vilela F, Vieira A, Kons R, Padilha UC, Vieira DCL, Bottaro M (2026)

Bilateral Deficit Using a Climbing-Specific Fingerboard Test and Its Association with Sport Climbers’ Performance. Sports, 14(7):276. doi:10.3390/sports14070276
n = 15 climbers. A 20 mm fingerboard on a Progressor 200: test–retest ICC 0.93–0.97 (bilateral and unilateral).
Takeaway: independent test–retest data for a Progressor 200 no-hang fingerboard test.

Pérez-Cordero J, Jerez-Mayorga D, Rodríguez-Perea Á, Soto García D (2025)

Reliability of finger strength assessment methods in climbing: a systematic review. Frontiers in Sports and Active Living, 7:1650198. doi:10.3389/fspor.2025.1650198
Systematic review of 15 studies (747 participants). Identifies the Progressor as the climbing-specific dynamometer “widely used among advanced climbers” and confirms Labott et al. as its published reliability and validity evidence.

Validation studies — clinical and sports science

Merry K, Napier C, Chung V, Hannigan BC, MacPherson M, Menon C, Scott A (2021)

The Validity and Reliability of Two Commercially Available Load Sensors for Clinical Strength Assessment. Sensors, 21(24):8399. doi:10.3390/s21248399
Bench test vs an Instron ElectroPuls E10000 across 10–1471 N. Test–retest ICC(3,1) 0.999–1.000; validity vs Instron ICC(3,1) ≥ 0.999, Pearson ρ ≥ 0.999; error below 0.9 N at 10–100 N and 5.1–7.3 N at 1000–1471 N; hysteresis mean 0.7%.
Takeaway: device-level accuracy — the sensor tracks a materials-testing reference to within the reference’s own error. (A bench study, not a patient study.)

Center JA, Willy RW, Elias ARC, Mischke JJ (2024)

Validity of a commercially available load cell dynamometer in measuring isometric knee extension torque in patients with knee disorders. Physical Therapy in Sport, 67:167–173. doi:10.1016/j.ptsp.2024.04.008
n = 31 patients with unilateral knee disorders. Agreement with isokinetic dynamometry for peak knee-extension torque: k = 0.91 (uninvolved) and 0.98 (involved limb), versus 0.84 and 0.93 for hand-held dynamometry. Limb-symmetry index agreement k = 0.89; sensitivity 96–100% and specificity 100% at the 70% cut-off, with better discrimination than hand-held dynamometry at the 90% cut-off.
Takeaway: in a real patient cohort the Progressor matched isokinetic testing at least as well as hand-held dynamometry.

Turner JA, Buonchristiani NA, Hartshorne MT, Noel S, Padua DA (2025)

Validity and Reliability of a Low-Cost and Portable Option for Measuring Peak and Explosive Isometric Quadriceps and Hamstrings Strength. Journal of Science in Sport and Exercise. doi:10.1007/s42978-025-00353-9
n = 28 healthy adults. Peak knee-extension and flexion torque vs isokinetic dynamometry: validity ICC 0.80–0.88; test–retest ICC 0.94–0.95 (extension) and 0.80–0.82 (flexion). Rate of torque development from the 80 Hz signal was not interchangeable with the isokinetic device.
Takeaway: good validity and excellent reliability for peak torque; treat rate-of-force-development values as device-specific.

Jacques D, Bazett-Jones DM, Garcia MC, Murray AM, Norte GE, Batista NP (2025)

Reliability and Validity of Low-Cost Tension Devices for Measuring Hip and Thigh Strength. Measurement in Physical Education and Exercise Science, 29(4):434–442. doi:10.1080/1091367X.2025.2495001
n = 19 healthy adults. Hip extension, abduction, internal and external rotation, and knee extension vs hand-held dynamometry: intra-rater reliability ICC 0.929–0.989; validity r 0.89–0.98 with no meaningful Bland–Altman bias.
Takeaway: excellent reliability and no systematic bias against hand-held dynamometry across five hip and thigh actions.

Sran A, Leake E, Yott S, Brandon JE, Power GA, Brandon SCE (2026)

Investigation of a Low-cost Portable Inline Dynamometer for Measuring Isometric Knee Extensor Strength. International Journal of Sports Physical Therapy, 21(7). doi:10.26603/001c.163287
n = 14. Progressor vs a wired 1000 Hz laboratory load cell: r = 1.0, ICC 0.94–1.00. Vs an isokinetic dynamometer the agreement depended on knee angle (ICC 0.37–0.80) with wide individual limits, attributed to the test set-up and lever arm rather than the sensor.
Takeaway: the sensor itself matches a wired laboratory load cell almost perfectly; set-up matters when comparing to isokinetic values.

Hickey JT, Lennon C, Gillick M, Sweeney L (2025)

Measuring eccentric hip adductor strength during the Copenhagen adduction exercise: a proof-of-concept and test re-test reliability study. Physical Therapy in Sport, 73:34–38. doi:10.1016/j.ptsp.2025.03.001
n = 20 field-sport athletes, two sessions. Test–retest ICC 0.84–0.87, SEM 3.3–3.6%.
Takeaway: good test–retest reliability for a novel eccentric hip-adductor field test built around the Progressor.

Studies using the Progressor as their measurement instrument

  • Cortés-Roco G et al. (2026). Plyometric training in boulder climbers. Sports, 14(8):313. doi:10.3390/sports14080313
  • McKellar BJ, Burr JF (2026). Blood flow restriction training and climbing performance. European Journal of Sport Science, 26(3). doi:10.1002/ejsc.70146
  • Iruretagoiena X et al. (2023). Ultrasound of finger flexor A2 pulley ruptures (calibrated tendon loading). Frontiers in Bioengineering and Biotechnology. doi:10.3389/fbioe.2023.1123857
  • Stock MS et al. (2026). Tracking muscle quality after ACL reconstruction. Journal of Functional Morphology and Kinesiology, 11(2):200. doi:10.3390/jfmk11020200
  • Dorca AJ et al. (2026). Shoulder function after latissimus dorsi breast reconstruction. Journal of Plastic, Reconstructive & Aesthetic Surgery, 122:28–37. doi:10.1016/j.bjps.2026.08.025
  • Ponce-Fuentes F et al. (2026). Exercise-induced hypoalgesia in rotator cuff repair rehabilitation. Translational Sports Medicine. doi:10.1155/tsm2/4698970
  • Micke F et al. (2026). Block vs traditional periodization in high-intensity functional training. Frontiers in Physiology. doi:10.3389/fphys.2026.1810477
  • Shelton AD et al. (2025). Hip abductor fatigue and gait instability in older adults. Journal of Neurophysiology. doi:10.1152/jn.00465.2024
  • Neltner TJ et al. (2025). Fatigability and neuromuscular responses to failure. Current Research in Physiology. doi:10.1016/j.crphys.2025.100149
  • Merry K et al. (2024). The PhysViz system for Achilles tendinopathy — the Progressor was selected after head-to-head testing of 11 load sensors. JMIR Human Factors, 11:e57873. doi:10.2196/57873
  • Montenegro GCS et al. (2026). Muscle oxygenation and perceived exertion in repetitive work. Revista Brasileira de Medicina do Trabalho. doi:10.47626/1679-4435-2026-1521
  • Darr AJ et al. (2026). RCT protocol, exercise treatment after mild traumatic brain injury. BMJ Open Sport & Exercise Medicine. doi:10.1136/bmjsem-2025-003065
  • Lanuza F et al. (2025). RCT protocol, plant-based foods and cardiometabolic risk. BMJ Open. doi:10.1136/bmjopen-2025-105461
  • Treinen Swake T et al. (2026). Novel isometric testing of frontal-plane hip strength in collegiate cross-country runners (Progressor 500 with a fixed strap). Sports Medicine International Open. doi:10.1055/a-2940-7121
  • Peterson (c. 2025). The test-retest reliability of a low-cost, portable tension dynamometer (Progressor 150, knee flexion and extension). UNC Office for Undergraduate Research abstract. our.unc.edu
  • Ordonez K, Scibek EP, Grimes JK (2025). Reliability and validity of Tindeq in measuring shoulder external rotator strength (poster). Sacred Heart University Academic Festival. digitalcommons.sacredheart.edu

What the research does not yet show

  • No published study has validated the Progressor 500 specifically; bench accuracy has been demonstrated up to about 1471 N (150 kg).
  • Rate of force development derived from the 80 Hz signal is device-specific and not interchangeable with high-sampling-rate laboratory dynamometers.
  • Climbing-specific validity against a force plate rests on one study (Labott et al., n = 25); finger-flexor tests have test–retest data but no independent force-plate comparison yet.
  • Only one validation study used a patient population (Center et al., knee disorders).

Last reviewed October 2026. Know of a study we’ve missed? Email or****@****eq.com.

Choosing an instrument

Most of the studies above used the Progressor 200 (or its predecessor, the 150); the Progressor 500 adds the range needed for whole-body and lower-limb tests. See Progressor 200 vs 500, or start with our explainer on Bluetooth dynamometers for measuring human strength.